Assist 4D probe; figure out the probe by way of the ID code; change the 4D push sign by using the  electrical relay to produce the 4D probe get the job done normally. 4.2.two Primary board The principle board can be divided into two parts: front-finish of the principle board and back again-conclusion of the most crucial board.

The user will not be permitted to open up the addresses and panel from the program, neither NOTE: gadget disassemble is permitted. To ensure the procedure performance and protection, only Mindray engineers or engineers licensed by Mindray can complete upkeep.

Get an entire and prolonged look at in the anatomical framework by panoramic imaging coupled with velocity indicator and ahead/backward scan means generating scanning easier, smoother plus more controllable.

Untuk memenuhi janji menyediakan perawatan kesehatan berkualitas tinggi dan efektif, Mindray terus mengembangkan kemajuan teknologi medis untuk meningkatkan kualitas perawatan pasien. Dilengkapi dengan PW Doppler dan kemampuan 3D/4D, DP-50Exp menawarkan lebih dari yang Anda harapkan.

Minimize off cable ties connecting the leading device about the Exhibit by diagonal chopping pliers, and then pull out the corresponding cable plugs.

Mindray’s new technology of black & white ultrasound procedure, DP-fifty Specialist, is equipped with advanced imaging technologies allowing for further penetration and speedier graphic acquisition with bigger resolution.

S. ought to deliver the Office using an e-mail address if they have 1. The e-mail delivered may very well be employed for Formal conversation With all the licensee. Nevertheless electronic mail addresses are community history. If you do not would like to supply a personal deal with, you should supply the Office using an electronic mail deal with which may be produced available to the general public. Make sure you see our Chapter 455 website page to ascertain For anyone who is influenced by this alteration.

Find out far better diagnostic details via prolonged check out on the anatomical framework on all convex probes.

Assessing people in nowadays’s demanding atmosphere involves modern technology that permits clinicians to generate quick and self-confident diagnoses. The flexibility of the DP-50 expands your medical utility when assisting to enhance individual care.

five.two.3 Geometric positioning accuracy Longitudinal geometric positioning accuracy  Examination Move: Do changes as how in screening the maximum depth. File the space by 20mm Each and every segment to the longitudinal targets line using the measurement caliper; Select the worth with the greatest error (to 20mm), calculate the accuracy utilizing the components down below The measurement caliper ought to be positioned with the higher fringe of the focus on, NOTE:... Web page 79 Transverse geometric positioning precision  Take a look at Move: Cover the scan surface in the phantom with water or couple gel, gently contact the probe Along with the scan surface Modify the depth, making the transverse targets to become exhibited from the graphic. Modify the focus stage being posited beside the transverse targets (the conventional will not be distinct) Modify parameters like attain, TGC, building Each individual transverse targets being Plainly displayed. Web site 80 five.5.two.four Blackout Spot Test Action: Deal with the scan floor from the phantom with h2o or few gel, Carefully Make contact with the probe with the scan surface Adjust the depth in a decrease benefit, and set the main target at the nearest spot to the scan surface. Decrease the value of parameters like AP, Acquire until finally the track record sound just may be seen. Webpage 81: Software program Upgrade And Upkeep

The DP-50 Expert is an innovative, superior-functionality Ultrasound Technique engineered to assist you to have the insights you would like. The Program’s powerful engine and superior-pace processing present fantastic graphic quality and a complete facts administration Resolution.

Find out better diagnostic information and facts through extended see of the anatomical structure on all linear probes

Overall performance Test five.five.1 Exam Course of action Resolution Detection depth Geometric Placement Precision Blackout spot check History the exam images 5.five.two Take a look at Information The image utilized Here's only for reference, persist with the impression effect in the true Take note: circumstance. Dp50rb login Prerequisites: Screen: established the distinction and brightness on the medical software benefit (or even the default standing) Procedure environment: dark area, simulating the scientific software surroundings. Web site seventy six In situation the transverse resolution screening targets are horizontally shown, report the negligible length of two targets that can be Obviously acknowledged. Repeat the operation previously mentioned for that transverse resolution screening targets at other depths. As demonstrated in determine underneath. Axial resolution ... Site seventy seven When utilizing the convex probe, hold the transverse resolution screening targets to NOTE: be displayed near the midline. When using a linear probe with steer function, don't turn on the steer function when perform the transverse resolution test. Zoom within the location where by the targets Found if important. The diameter with the goal level at a certain depth is equal for the transverse resolution in the depth. Web page 78 five.

Your Software for deeper biopsy: letting changes for the scan line to realize greater visibility with the needle, nerves and little vessels, guaranteeing the continuity of all knowledge.

Take note: Make certain that you could pull out and drive in the IO board sockets vertically when pulling out and assembling the IO board assembly on the placement with the deal with. Take away panhead screws with washers M3X6 (four) utilized to protected the IO board, and after that you might acquire out the IO board. Web site 116 Eliminate panhead screws with washers (2 M3X6) which can be mounted on the leading device, take out the higher baffle and then pull out two cables connecting the key board to the power module assembly.